Powerheart AED G3 Features
| |
 |
 |
|
Critical
to saving victims of sudden cardiac arrest is early access to defibrillation.
The easy-to-use features of the Powerheart AED G3 allow virtually anyone
to defibrillate cardiac arrest victims quickly and effectively, dramatically
improving the chances of survival from a cardiac arrest.
- Opening the lid of the Powerheart AED G3 turns the AED on and readies
it for use.
- More instructive voice and text prompts guide the user through every
step of the rescue process.
- Simple, one-button operation eliminates rescuer confusion and uncertainty.
- Patented RescueReady® technology performs daily, weekly, and monthly
self-tests on the battery, electrodes (presence and function), internal
electronics, and software.
- Daily automated self-test includes: battery, electrodes (presence
and function), internal electronics, SHOCK/CONTINUE button, and software.
- Weekly automated self-test includes: battery, electrodes (presence
and function), internal electronics, partial energy charge, SHOCK/CONTINUE
button, and software.
- Monthly automated self-test includes: battery, electrodes (presence
and function), internal electronics, full energy charge cycle, SHOCK/CONTINUE
button, and software.
- Powered by an Intellisense® lithium battery.
- Industry's first four year full replacement battery warranty.
- SmartGauge® indicator and daily battery capacity test monitors
battery status and alerts the user when the battery needs to be replaced.
- STAR® Biphasic Waveform technology delivers customized, escalating
variable energy based on the patients needs, to assure rapid defibrillation
success.
- Escalating
Variable Energy (VE) 105J-360J.
- Pediatric capability.
- Five programmable energy protocols.
- Pacemaker pulse detection.
- Advises for shocks only when appropriate, and continues to monitor
the patient even after AED makes shock decision and charges (the non-committed
feature will not allow shock delivery if the rhythm changes to non-shockable).
- Detects and advises defibrillation therapy for VF, VT and SVT as recommended
by the American Heart Association (AHA).
- Programmable variable energy protocol options. Optional: only AED
to offer programmable detection rates for VF, VT and SVT (pre-programmable
by a Medical Director using MDLink software).